Advanced Instruments Inc.
2855 Metropolitan Place, Pomona, CA 91767 USA Tel: 909-392-6900, Fax: 909-392-3665, e-mail: [email protected], www.aii2.com
Owner’s Manual
GPR-3100
Oxygen Purity Analyzer
Advanced Sensor Technology
24 Month Operating Life in 100% Oxygen
Inherent Accuracy 0.1%, Resolution 00.01%
Auto Ranging & Suppressed Range Capability
Temperature Controlled & Barometric Compensationsy
Vista de página 0
1 2 3 4 5 6 ... 44 45

Resumo do Conteúdo

Página 1 - Owner’s Manual

Advanced Instruments Inc. 2855 Metropolitan Place, Pomona, CA 91767 USA ♦ Tel: 909-392-6900, Fax: 909-392-3665, e-mail: info@aii2.

Página 2

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 3 Sample System: The GPR-3100 is supplied with sta

Página 3

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 4 throttle valve on the pump’s inlet side may be n

Página 4

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 5 Calibration and Accuracy Single Point Calibratio

Página 5

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 6 Graph C illustrates these 'worse case'

Página 6

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 7 Installation The GPR-3100 Oxygen Analyzer consis

Página 7

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 8 Mounting the Analyzer: The GPR-3100’s 10.8”W x 7

Página 8

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 9 Gas Connections: The GPR-3100 is designed for po

Página 9

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 10 Procedure: Caution: Do not change the factory s

Página 10 - 5 Operation

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 11 Output Connections As illustrated above the sen

Página 11

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 12 Power Failure Alarm A dry contact rated at 30VD

Página 12

Advanced Instruments Inc. Table of Contents Introduction 1 Quality Control Certification 2 Safety 3 Features & Specifications 4 Operation

Página 13

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 13 Installing the Oxygen Sensor The GPR-3100 is eq

Página 14

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 14 Installing Span Gas Caution: Do not contaminat

Página 15

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 15 Establishing Power to the Electronics: Once the

Página 16

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 16 Operating the Menu Driven Controls – The appen

Página 17

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 17 Range Selection: The analyzer display defaults

Página 18

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 18 4. Press the ENTER key when F-C starts flashin

Página 19

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 19 Calibration - General Guideline Zeroing the ana

Página 20

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 20 consume the oxygen that has dissolved into the

Página 21

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 21 span calibration again for up to 3 months under

Página 22

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 22 16. The analyzer returns to the SAMPLE mode aft

Página 23

Advanced Instruments Inc. 1 Introduction Thank You For Your Business Your new oxygen analyzer is a precision piece of equipment designed to g

Página 24

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 23 To avoid damaging the sensor or sample system o

Página 25

Advanced Instruments Inc. 6 Maintenance GPR-3100 Rev 7/2004 6

Página 26

Advanced Instruments Inc. 7 Spare Parts Recommended spare parts for the GPR-3100 % High Purity O2 Analyzer include: Item No. DescriptionGPR-11

Página 27

Advanced Instruments Inc. 8 Troubleshooting Symptom Possible Cause Recommended Action Reading does not reflect expected values Sensor was not c

Página 28

Advanced Instruments Inc. 9 Warranty What is covered: Any defect in material and workmanship from normal use in accordance with the Owner’s Manu

Página 29

Advanced Instruments Inc. 9 Warranty What is not covered: This warranty does not cover installation; defects resulting from accidents; damage wh

Página 30

Advanced Instruments Inc. 10 Material Safety Data Sheet (MSDS) Product Identification Product Name Oxygen Sensor Models CAD, GPR, PSR, SAF, 670

Página 31

Advanced Instruments Inc. 10 Material Safety Data Sheet (MSDS) Fire and Explosion Data Flash and Fire Points Not applicable Flammable Limits N

Página 32

Advanced Instruments Inc. 10 Material Safety Data Sheet (MSDS) Symptoms Eye contact - burning sensation. Skin contact - soapy slick feeling. Me

Página 33

Advanced Instrtuments Inc.5 Operation - Appendix A* MAIN MENU Sample Sample Span Zero Alarm SystemAuto Range 90 t

Página 34

Advanced Instruments Inc. 2 Quality Control Certification Date: Customer: Order No.: Pass Mo

Página 35

Advanced Instrtuments Inc.5 Operation - Appendix A* MAIN MENU Sample Sample Span Zero Alarm SystemAuto Range 90 t

Página 36

Advanced Instrtuments Inc.5 Operation - Appendix A* MAIN MENU Sample Sample Span Zero Alarm SystemAuto Range 90 t

Página 37

Advanced Instrtuments Inc.5 Operation - Appendix A* MAIN MENU Sample Sample Span Zero Alarm SystemAuto Range 90 t

Página 38

Advanced Instrtuments Inc.5 Operation - Appendix A* MAIN MENU Sample Sample Span Zero Alarm SystemAuto Range 90 t

Página 39 - 99.55 %

Advanced Instrtuments Inc.5 Operation - Appendix A* MAIN MENU Sample Sample Span Zero Alarm SystemAuto Range 90 t

Página 40 - 99.55 % 99.55 %

Advanced Instrtuments Inc.5 Operation - Appendix A* MAIN MENU Sample Sample Span Zero Alarm SystemAuto Range 90 t

Página 41 - 00.05 %

Advanced Instruments Inc. 3 Safety This section summarizes the essential generic precautions applicable to all analyzers. Additional precautions

Página 42 - 5 Operation - Appendix A

Advanced Instruments Inc. 3 Safety Serviceability: Except for replacing the oxygen sensor, there are no parts inside the analyzer for the operato

Página 43

Technical Specifications Accuracy: < 1% of FS range at constant conditions; inherent accu-racy of 0.1% following calibration with 95-100% oxygen

Página 44

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 1 Principle of Operation The GPR-3100 oxygen analy

Página 45

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 2 With the sensor and sample gas lines temperature